General DPI Documentation

These can be found at \\opus.cvs.rochester.edu\aplab\DPI_Wiki_RIT

Screaming Motor Syndrome

If a motor is continuously screaming, it is likely that some of the pivot screws just need to be tightened. For example, if one of the motors if noisy, but you can track successfully by holding the mirror stationary, then this may be the fix for you. Warren's video instructions for repairing the motor are available on OPUS as a .exe file (but is a video):

\\opus.cvs.rochester.edu\aplab\DPI_Wiki_RIT\Warren_Movies\mirror-motor-repair.exe

Warren Rochester Visit Notes

These can be found at \\opus.cvs.rochester.edu\aplab\DPI_Wiki_RIT\WWW_2018_visit_materials

He also pointed us to number of videos, which can be found at \\opus.cvs.rochester.edu\aplab\DPI_Wiki_RIT\Warren_Movies

A page of troubleshooting notes

Procedure for Focus

"Procedure:

  1. Set up static model eye.
  2. Focus Lock Up (on).
  3. Lock on model eye. Move Lens 8 slightly (first large lens) to be near 0.0.
  4. Perform focus detector adjustment to get equal sine waves on test points 1 and 2 on Card 8.
  5. Turn off Focus Lock and see if Lens 8 stays at or near 0.0.

If this doesn’t work, there may be another problem." -- WWW 6/5/18

Noise reports

DPI baseline reports (noise levels, resolution, dynamics) and maintenance history are located on this page: DPIBaseline

Galvo instructions and calibration report

Instructions for using the galvo with the DPI artificial eye. Calibration of the galvo in its current configuration (attached to the correct galvo driver, with the correct gain setting) was conducted on 10/6/20 by Ben Moon. Calibration data, analysis code, and other documentation is here on Opus: opus.cvs.rochester.edu\aplab\DPI_GalvoCalibration.

Instructions for DPI Galvo
Calibration Report (2020-10-22, BM)

Equipment-DPIEyeTracker (last edited 2020-10-26 16:14:37 by BenjaminMoon)